15x15 Ethanol Sensor 1000 ppm C Package 110-205 - 3SP_Ethanol_1000 C
• Measurement Range: 0 to 1000 ppm
• Resolution: < 5 ppb (instrumentation dependent)
• Repeatability: < +/- 3 % of reading
• Response Time: < 180 seconds typical
• Sensitivity: 15 +/- 6 nA/ppm
• Small Size with Low Profile (20 x 20 x 3 mm)
• Long Life (10 years expected life)
• Fast Response
- Quantity:
- - +
Product Specification
3SP_Ethanol_1000 C
15x15 Ethanol Sensor 1000 ppm C Package 110-205
3SP_Ethanol_1000 C Breath Alcohol Sensor Description
SPEC's printed sensors offer the performance of the best quality electrochemical sensors at a fraction of the price. SPEC's printed sensors are also ultra-thin, offering easy integration into wireless, portable, and networked solutions. These sensors are ideal for health, environmental, industrial and residential monitoring, because of their high performance, low cost and small size.
• Small Size with Low Profile (20x20x3 mm)
• Long Life (10 years expected life)
• Fast Response
• Low Power Consumption
• Individually Calibrated (NIST Traceable)
• ROHS Compliant
3SP_Ethanol_1000 C Breath Alcohol Sensor Applications
• Law Enforcement Breathalyzers
• Evidential Breath Alcohol Testing
• Portable Breath Alcohol Tester
• Personal Breathalyzers
3SP_Ethanol_1000 C Breath Alcohol Sensor Specifications
Measurement Range
0 to 1000 ppm
Resolution
< 5 ppb (instrumentation dependent)
Repeatability
< +/- 3 % of reading
Response Time
< 180 seconds typical
Sensitivity
15 +/- 6 nA/ppm
Expected Operating Life
> 5 years (10 years @ 23+/-3C; 40+/-10% RH)
Operating Temperature Range
-30 to +55 C (-20 to 40 C recommended))
Operating Humidity Range - non-condensing
0 to 100% RH (15 to 95% recommended)
Power Consumption
10 to 50 uW (circuit & ambient Ethanol dependent)
3SP_Ethanol_1000 C Breath Alcohol Sensor Cross Sensitivity
Gas/Vapor
Concentration
Typical Response PPM Ethanol
Methane
500 ppm
< 5
Ammonia
100 ppm
< 5
Nitrogen Dioxide
10 ppm
< 5
Hydrogen Sulfide
25 ppm
63.2
Carbon Monoxide
400 ppm
251.4
Ozone
5 ppm
< 5
Sulfur Dioxide
20 ppm
11.6
Nitric Oxide (NO)
50 ppm
54.5
Chlorine
10 ppm
-14.3
n-Heptane
500 ppm
-12.6
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Electrochemical carbon monoxide sensor CO
-
Carbon Monoxide Sensor (CO Sensor)
-
Hydrogen Cyanide Sensor (HCN Sensor)
-
Phosphine Sensor (PH3 Sensor)
-
Nitric Oxide Sensor (NO Sensor)
-
Nitrogen Dioxide Sensor (NO2 Sensor)
-
Infrared Carbon Dioxide (CO2) Sensor Pyroelectric Detector
-
Ethylene Oxide Sensor (ETO Sensor)
-
Individual Sensor Board (ISB) Alphasense B4 4-Electrode Gas Sensors
-
Hydrogen Chloride Sensor (HCL Sensor)