15x15 Carbon Monoxide Sensor 1000 ppm Package 110-109 - 3SP_CO_1000 C
• Measurement Range: 0 to 1,000 ppm
• Detection Limit: 0.5 ppm
• Resolution: < 100 ppb (instrumentation dependent)
• Repeatability: < ± 2 % of reading
• Response Time - T(90): < 30 seconds (15 seconds typical)
• Sensitivity: 4.75 ± 2.75 nA/ppm
• Power Consumption: 10 to 50 uW (circuit & ambient CO dependent)
- Quantity:
- - +
Product Specification
3SP_CO_1000 C
15x15 CO Sensor 1000 ppm Package 110-109
SPEC Sensors’ Screen Printed ElectroChemical sensor technology (SPEC Sensor™) revolutionizes the current state of the art, enabling new applications in consumer and industrial safety monitoring. SPEC’s printed sensors offer the performance of the best quality electrochemical sensors at a fraction of the price. SPEC’s printed sensors are also ultra-thin, offering easy integration into wireless, portable, and networked solutions. These sensors are ideal for health, environmental, industrial and residential monitoring, because of their high performance, low cost and small size.
3SP_CO_1000 C Carbon Monoxide Sensor (CO Sensor) Benefits
• Small Size with Low Profile (20x20x3.8 mm)
• Long Life (10 years expected life)
• Fast Response (< 15 seconds)
• Robust (passes 5000 ppm overload)
• Low Power (0 mW @ 0 mV bias)
• Individually Calibrated (NIST Traceable)
• ROHS Compliant
3SP_CO_1000 C Carbon Monoxide Sensor (CO Sensor) Applications
• Residential and Commercial CO Monitoring
• Industrial CO Monitors
• Ventilation Control
• RV and Marine CO Monitoring
• Indoor Air Quality
• Outdoor Air Quality
3SP_CO_1000 C Carbon Monoxide Sensor (CO Sensor) Specifications
Measurement Range |
0 to 1,000 ppm |
Detection Limit |
0.5 ppm |
Resolution |
< 100 ppb (instrumentation dependent) |
Repeatability |
< ± 2 % of reading |
Response Time - T(90) |
< 30 seconds (15 seconds typical) |
Sensitivity |
4.75 ± 2.75 nA/ppm |
Overload |
Passes EN20291-1 Sec. 5.3.6 5,000 ppm overload |
Expected Operating Life |
> 5 years (10 years @ 23 ± 3 °C; 40 ± 10% RH) |
Operating Temperature Range |
-30 to 55 °C (-20 to 40 °C continuous recommended) |
Operating Humidity Range- non-condensing |
15 to 95% recommended continuous 0 to >95% RH - intermittent |
Operating Bias |
0 to 5 mV |
Power Consumption |
10 to 50 uW (circuit & ambient CO dependent) |
3SP_CO_1000 C Carbon Monoxide Sensor (CO Sensor) Cross Sensitivity
Most chemical sensors exhibit some cross-sensitivity to other gases. The following table lists the relative response of common potential interfering gases, and the concentration at which the data was gathered.
Gas/Vapor |
Concentration |
Typical Response PPM CO |
Carbon Dioxide |
5,000 ppm |
< 1 |
Hydrogen |
100 ppm |
17 |
Methane |
3,000 ppm |
< 1 |
Ammonia |
100 ppm |
< 1 |
Nitrogen Dioxide |
10 ppm |
< 1 |
Hydrogen Sulfide |
25 ppm |
< 1 |
Carbon Monoxide |
400 ppm |
400 |
Ozone |
5 ppm |
< 1 |
Sulfur Dioxide |
20 ppm |
< 1 |
Chlorine |
10 ppm |
< 1 |
n-Heptane |
500 ppm |
< 1 |
Toluene |
200 ppm |
< 1 |
Isopropyl Alcohol |
200 ppm |
1.3 |
Acetone |
200 ppm |
< 1 |
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Electrochemical carbon monoxide sensor CO
-
Carbon Monoxide Sensor (CO Sensor)
-
Hydrogen Cyanide Sensor (HCN Sensor)
-
Phosphine Sensor (PH3 Sensor)
-
Nitric Oxide Sensor (NO Sensor)
-
Nitrogen Dioxide Sensor (NO2 Sensor)
-
Infrared Carbon Dioxide (CO2) Sensor Pyroelectric Detector
-
Ethylene Oxide Sensor (ETO Sensor)
-
Individual Sensor Board (ISB) Alphasense B4 4-Electrode Gas Sensors
-
Hydrogen Chloride Sensor (HCL Sensor)