Membrane Covered Optical CO2 Sensor for Monitoring - OPTCO2

  • Membrane Covered Optical CO2 Sensor for Monitoring
  • Membrane Covered Optical CO2 Sensor for Monitoring
Mouse over image to zoom

• Measuring ranges: 15 mg/l, 50 mg/l, 340 mg/l and others on request (please select one)​
• Dimensions​: 225 mm length, 33 mm diameter​
• Housing material​: Titanium​
• Power consumption​: < 0.5 W​
• Warm up time​: < 12 seconds, < 2 minutes for full accuracy​
• Maximum deployment depth​: 1,000 m/1,000 dbar​
• Available outputs​: 4...20 mA, 0...5 V DC or digital output RS485​

  • Model Number: OPTCO2
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

Membrane covered optical CO2 sensor for monitoring
Online determination of the dissolved carbon dioxide in water


The measurement of dissolved carbon dioxide (excluding HCO3- and CO32-) in water is one of the most important parameters for water monitoring. AMT GmbH has developed now a new membrane covered optical CO2 sensor with titanium housing and other improved characteristics. So the new AMT-CO2 sensor is designed for depths of up to 1,000 m, has an analogue or digital output, a reduced current uptake and includes an automatic internal temperature compensation. The sensor is designed to be interfaced externally to CTD probe systems, to own measuring systems or as stand alone sensor.

Main features of the Carbon Dioxide Sensor:
- submersible direct reading sensor for natural and industrial waters and sea water
- detects CO2 and total inorganic carbon when measuring additionally pH
- no signal interferences to silicate, phosphate, HCO3- and CO32-
- built-in reference measurement for superior stability
- calibration coefficients stored internally
- for depths of up to 1,000 m
- less than 0.5 W power consumption
- with analogue or digital output (please select)
- with internal temperature compensation
- compensation of humidity and other influences
- heated sensor head to avoid condensation

Technical data of the membrane covered optical carbon dioxide sensor:
Carbon Dioxide:
Measuring ranges
15 mg/l, 50 mg/l, 340 mg/l and others on request (please select one)
Accuracy
0...5 mg/l CO2: ± 0.06 mg/l
5...15 mg/l CO2: ± 2 % of the measuring value
15...50 mg/l CO2: ± 3.5 % of the measuring value
50...80 mg/l CO2: ± 1.5 mg/l CO2
80...340 mg/l CO2: ± 6 mg/l CO2
General:
Dimensions
225 mm length, 33 mm diameter
Housing material
Titanium
Connector
Subconn MCBH4M
Power supply
12-30 V DC when using the digital or voltage output or
20-30 V DC when using the current output
Power consumption
< 0.5 W
Warm up time
< 12 seconds, < 2 minutes for full accuracy
Maximum deployment depth
1,000 m/1,000 dbar
Available outputs
4...20 mA, 0...5 V DC or digital output RS485
Temperature range for storage and measurement
0...60°C
Internal temperature compensation
Compensation of humidity and other influences

If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service