Remote Particle Counter - P235
• Measures 0.5 μm and 5.0 μm
• 0.1 CFM (2.83 LPM) flow rate
• Measures 2 channels of simultaneous data
• Power: 9-30 VDC
• Long life laser diode
• Concentration limit of 2,000,000 per ft³
• Operating Temperature: 5 to 35°C
• Operating Humidity: 20% to 95%RH non-condensing
- Quantity:
- - +
Product Specification
Remote Particle Counter
Model P235
The P235 remote particle counter measures 0.5 µm and 5.0 µm with a flow rate of 0.1 CFM (2.83 LPM). Integration is easy into a real-time monitoring or building management system via Modbus™ RTU or Modbus™ ASCII. The P235 is the smallest remote particle counter in the market and has easy mounting options. The P235 complies with ISO 21501-4 and includes a 2 year limited warranty.
P235 Remote Particle Counter Features & Benefits
• Measures 0.5 µm and 5.0 µm
• 0.1 CFM (2.83 LPM) flow rate
• Measures 2 channels of simultaneous data
• Long life laser diode
• Modbus™ RTU or Modbus™ ASCII outputs
• Concentration limit of 2,000,000 per ft³
• Complies with ISO 21501-4 and JIS B9921 standards
• Lightweight stainless steel enclosure
• Easy to clean and wipe down
• 2 year limited warranty
P235 Remote Particle Counter Specifications
Channel Sizes | 0.5 µm and 5.0 µm |
Counting Efficiency | 50% at 0.5 µm; 100% for particles > 0.75 µm (per JIS) |
Concentration Limits | 2,000,000 particles / ft3 at 5% coincidence loss |
Light Source | Long life laser diode |
Zero Count Level | <1 count / 5 minutes (per JIS B9921) |
Flow Rate | 0.1 CFM (2.83 LPM) |
Flow Control | Critical orifice |
Vacuum Requirements | External vacuum >15" (38.1 cm) of Hg |
Calibration | NIST traceable |
Sample Probe/Tubing | Isokinetic sampling probe |
Communication Interface | Modbus™ RTU or ASCII outputs (specified upon order) |
Power | 9-30 VDC |
Dimensions (L x W x H) | 1.8" x 3.7" x 1.5" (4.5 cm x 9.3 cm x 3.8 cm) does not include probes |
Weight | 0.62 lb (0.28 kg) |
Standards | ISO 21501-4 and JIS B9921 |
Warranty | 2 year limited warranty |
Operating Conditions | 5 to 35°C, 20% to 95%RH non-condensing |
Storage Conditions | -20 to 50°C, Up to 98%RH non-condensing |
Optional Accessories | Isokinetic sampling probe, purge filter, data cable, vacuum and sample tubing |
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
4-channel Remote Particle Counter
-
Remote Particle Counter
-
Remote Particle Counter
-
Remote Airborne Particle Counter 0.3 µm @ 0.1 CFM
-
Remote Airborne Particle Counter 0.5 µm @ 0.1 CFM
-
Remote Particle Counter 0.3 µm @ 0.1 CFM
-
Remote Particle Counter 0.5 µm @ 0.1 CFM
-
Handheld Particle Counter 0.3 µm @ 0.1 CFM
-
Handheld Particle Counter 0.3 µm @ 0.1 CFM
-
Handheld Particle Counter 0.5 µm @ 0.1 CFM