• Measuring range​: four measuring ranges with end value of 2, 20, 200, 2000 W/m2 , automatic shifting​
• Accuracy​: ≤ ± 5 % of the measuring range related to the standard of PTB Braunschweig​
• Memory​: the last measured value is stored​
• Analogue output​: 0.1 to 4.1 V, R i = 10 kΩ, jack 2.5 mm, signal inside​
• Degree of protection​: ​IP 65 with battery / connection box closed​
• Temperature range: ambient temperature 0 to 30°C​

  • Model Number: MUV 2.4 WR
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

Reference radiometer MUV 2.4 WR with DVGW and/or ÖNORM sensor

MUV 2.4 WR Reference Radiometer Description
The UV measuring device MUV2.4WR in connection with UV sensors is designed as a reference measuring unit according to DVGW/ÖNORM. It is set for the check-up of UV systems which are used for drinking water disinfection. Based on comparison measurements between field sensor and reference sensor, it is possible to figure out, whether the field sensor must be re-calibrated or exchanged.

A closed sturdy metal housing protects the unit against external hazard and ensures the IP65 degree of protection. The device is run by rechargeable batteries, equipped with a LCD display with backlight, automatic measuring range changer and can be controlled easily by only 3 buttons. Battery recharge and power supply can be done by a wall power supply and it is possible to connect a printer or computer. The measuring device comes in a transport case.

MUV 2.4 WR Reference Radiometer Operators
System designer and system builder
Technicians in water plants
Health offices
Laboratories

MUV 2.4 WR Reference Radiometer Technical data

Housing
handheld aluminum housing, colored matt black
Size L x H x D
105 x 230 x 35 mm
Weight
790 g incl. sensor and rechargeable battery
Power supply
internal accumulator
2 x 1.2 V / 2 Ah, size AA (Mignon), recharge time approx. 4 h at 0.5 A, operating time 10 h (during
battery recharge the control LED in the connection box is lighting)
Mains adaptor
110 - 230 V AC/ 9 - 12 V DC 5 W, mains plug EU (Art.-No.: 321 11000 0000)
On request with additional adaptors for US or UK (Art.-No.: 321 11005 0000)
Display
LCD, 2 x 16 characters with backlight function
Measuring range
four measuring ranges with end value of 2, 20, 200, 2000 W/m2 , automatic shifting
Accuracy
≤ ± 5 % of the measuring range related to the standard of PTB Braunschweig
Memory
the last measured value is stored
Analogue output
0.1 to 4.1 V, R i = 10 kΩ, jack 2.5 mm, signal inside
Serial interface
9600 Baud, no parity, 1 stop bit, 8 data bits, GND/TxD/RxD
connection via 3.5 mm jack bush
Degree of protection
IP 65 with battery / connection box closed
Temperature range
ambient temperature 0 to 30°C
Recalibration
Latest after 12 month, according to DVGW W294 rule
Accuracy
Uncertainty to working standard: ≤ ± 1% for low pressure, ≤ ± 3% for medium pressure lamps
Uncertainty to internal reference: ≤ ± 1%
Uncertainty to primary standard DVGW/ÖNORM: ≤ ± 6.5%
Total uncertainty: ≤ ± 8.5% for low pressure, ≤ ± 10.5% for medium pressure lamps
Calibration protocol included in scope of delivery.