UV Radiometer 2.0 - GUVx-T1xGS2-I8LO

  • UV Radiometer 2.0
  • UV Radiometer 2.0
Mouse over image to zoom

- High Temperature Environment (under 200℃)
- Output of Display: Relay, 4-20mA Current
- Fixing method: Panel Insertion
- Power Consumption: Max 5VA
- Detection Power Range: 0~1000 ㎽/㎠
- Detection Range:
GUVA-T13GS2-I8LO: 230~395nm
GUVC-T11GS2-I8LO: 220~280nm

  • Model Number: GUVx-T1xGS2-I8LO
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

GUVx-T1xGS2-I8LO UV Radiometer 2.0 Description
 High Temperature Environment (under 200°C), Output of Display : Relay, 4-20mA Current

GUVx-T1xGS2-I8LO UV Radiometer 2.0 Constitution
 Display, Optical Fiber + LO Probe, Cable

GUVx-T1xGS2-I8LO UV Radiometer 2.0 Case Dimensions
Display
LO Probe Diameter (mm)
LO Probe Length (mm)
Optical Fiber Length (m)
Size
 Panel Cut Out
95 x 48 x 118 mm3
92 x 45 mm2 (-0,+0.6mm)
20 56 1


GUVx-T1xGS2-I8LO UV Radiometer 2.0 Absolute Maximum Ratings

Parameter
Value
Unit
Remark
Min.
Typ.
Max.
Operating Temperature


200 °C
Optical Fiber
-30

85
°C
LO Probe
0

50 °C
Display
Power Consumption


5
VA


GUVx-T1xGS2-I8LO UV Radiometer 2.0 Electro-Optical Characteristics (at 25 °C)

Parameter
Value
Unit
Remark
Min.
Typ.
Max
Detection Range
GUVA-T13GS2-I8LO
230
395 nm
GUVC-T11GS2-I8LO
220
280 nm

Detection Power Range
0
1000 mW/cm2