Oxygen Sensor - PSR-11-915-2i

  • Oxygen Sensor
Mouse over image to zoom

USA manufactured and medical certified electrochemical galvanic oxygen sensors with excellent accuracy and stability and individually tested prior to shipment.
Fresh sensors direct from manufacturing with extended lifetimes and long warranty.

  • Model Number: PSR-11-915-2i
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

    USA manufactured and medical certified electrochemical galvanic oxygen sensors with excellent accuracy and stability and individually tested prior to shipment.

    Fresh sensors direct from manufacturing with extended lifetimes and long warranty.

 

Technical Specifications

Measuring Range

0-100% O2

Signal Output (1)

45-65 mV

Response Time 90%

9 Seconds

Accuracy Full Scale (2)

±1%

Accuracy FS Over Operating Range (3)

±5%

Drift % Signal Month (2)

< 1%

Linearity (2)

±1%

Humidity Non-Condensing

0-99% RH

Temperature Coefficient

Compensated

Operating Temperature

0-45

Operating Pressure

Constant 5-30 PSIG

Flow Rate

.1-10 LPM

Expected Life (1)

60 Months

Storage (4)

0 to 40

Warranty (5)

18 Months

Electrical Connections

3 Ring PCB Gold Plated

Packaging

Polyetylene bag in 3" x 3" x 1.5" paper box

Recommended Storage

< 6 months

(1) In air @ 25and 1 atm. Expected life varies inversely with changes in O2 concentration, temperature, pressure.

(2) Under constant conditions. Calibrate as near sample conditions as possible. Calibrate with 95-100% oxygen for measurements > 40% oxygen.

(3) Once the signal output stabilizes following a step change in temperature.

(4) Recommended 0-40. Sensors may be stored at -10 to 55on an intermittent basis only.

(5) Ex-factory unless agreed to in writing. Under normal application conditions sensor is warranted to be free of defects in material and workmanship for the period specified above proved the sensor is properly installed and operated. Analytical Industries Inc. shall not be liable for buyer's negligence,misapplication, alteration or abuse. The sole remedy for a sensor determined to be defective by Analytical Industries Inc. is limited to replacing the defective sensor.



If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service