Mass Flowmeter for Gases - 4143

  • Mass Flowmeter for Gases
  • Mass Flowmeter for Gases
  • Mass Flowmeter for Gases
Mouse over image to zoom

• Fast and accurate
• Low pressure drop
• Gas Calibration: Air, O2, N2
• Inlet/Outlet Diameter: 0.375” (9.53 mm)
• Flow Measurement Range: 0.01-20 Std L/min
• Volume Measurement Range: 0.01 - 99.9 liters​
• Pressure Measurement​ Range​: 50-199 kPa absolute​
• Temperature Measurement​ Range​: 0-50°C​
• ​DC Power Input​: 7.5 VDC ±1.5 V, 300 mA max​

  • Model Number: 4143
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

4143 Mass Flowmeter PRODUCT DETAILS
The TSI Mass Flowmeter 4143 (LCD, 0.01-20L/min Air, O2 , N2 3/8" tube) can be used for a multitude of gas flow measurement applications. Whether measuring gas flows in a laboratory or manufacturing setting, TSI general purpose mass flowmeters provide accurate results with multiple data output options. The display versions come complete with accessories making setup and operation fast and convenient. The TSI mass flowmeters are specified as two percent of flow rate reading. One TSI mass flow meter covers the same range as three percent (or more) of full scale devices, with better accuracy at most points.

Four millisecond response ensures accuracy in fluctuating flows. This fast response time of the TSI mass flowmeter is ideal for closed-loop control systems and integrated volume measurements. Low pressure drop of the TSI mass flowmeter minimizes flow circuit back pressure and its impact on the system.

4143 Mass Flowmeter FEATURES AND BENEFITS
• Fast and accurate 
• Low pressure drop 

4143 Mass Flowmeter APPLICATIONS
• Medical 
• Industrial 
• Laboratory

4143 Mass Flowmeter SPECIFICATIONS – DIGITAL DISPLAY MODELS

Model
4143
Gas Calibration
Air, O2, N2
Inlet/Outlet Diameter
0.375” (9.53 mm)
Flow Measurement
Range
0.01-20 Std L/min
Accuracy — Air and O2
±2% of reading or 0.005 Std L/min, whichever is greater
Accuracy — N2
±3% of reading or 0.010 Std L/min, whichever is greater
Accuracy — Air and O2 mixture
N/A
Accuracy — N2O
N/A
Response
4 ms to 63% of full scale flow
LCD Display Units
L/min, Std L/min, cm 3 /min, Std cm 3 /min
Overall Dimensions
5” x 2” x 1.25” (127 mm x 49 mm x 32 mm)
Volume Measurement
Range
0.01 - 99.9 liters
Accuracy
±2% of reading (see Operator’s Manual for additional details) 
Pressure Measurement
Range
50-199 kPa absolute
Accuracy
±1 kPa
Response
<4 ms to 63% of final value for step charge
Temperature Measurement
Range
0-50°C
Accuracy
±1°C at flow greater than 1 Std L/min
Response
<75 ms to 63% of final value for step change
Outputs
Analog
0-10 VDC flow only, zero and span adjustable via RS232
Digital
RS232
DC Power Input
7.5 VDC ±1.5 V, 300 mA max

If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service