Generic Sensor Board for Oxygen Sensors by SENSORE Electronic - GSB
• Microcontroller based interface circuit
• Compatible O2-Sensor-types: SO-zz-xxx, SP-zz-xxx
• Typical current consumption 200mA @ 12Vdc
• Linear sensor output signal according to O2-concentration
• External voltage supply: 12Vdc
• Configurable sensor operation parameters
• Heater temperature control
- Quantity:
- - +
Product Specification
Generic Sensor Board
"GSB" (Generic Sensor Board) is a universal electronic module for all sensor types. It can be used both for controlling the amperometric and potentiometric sensors.
At this module, sensor temperature is precisely controlled to a constant temperature by microcontroller. The linearized output signal is available as a voltage signal (0-5V), current signal (4-20mA) or digital via RS232.
Electronic sensor interface for oxygen sensors
• Sensor interface board for direct integration in customer application
• Demonstration board for laboratory sensor evaluations
GSB - Generic Sensor Board Main Features
• Microcontroller based interface circuit
• Compatible O2-Sensor-types:
Standard: SO-zz-xxx (amperometric sensor type)
On request: SP-zz-xxx (ampero-potentiometric sensor type)
• External voltage supply: 12Vdc (possible range 6-25Vdc)
typical current consumption 200mA @ 12Vdc
• Sensor calibration
Single point calibration at specified O2-concentration
heater resistance calibration to ensure correct sensor temperature
Calibration values are stored in nonvolatile memory
possibility to purchase a factory calibrated set (GSB+sensor)
• Linear sensor output signal according to O2-concentration:
analog 0-5V (Vout)
analog 4-20mA (Iout)
digital via RS232
• Configurable sensor operation parameters
6 preinstalled settings for different sensor types
optional: customized parameters, e.g. :
▪ heatup time and heater operation mode
▪ sensor voltage
▪ Measurement range and corresponding output signals
• Heater temperature control:
4-wire heater resistance measurement
constant heater resistance control for temperature compensation
optional: constant heater voltage or constant power control
• Sensor connection:
connection via 6pol RAST 2.5
connection via screw terminal
optional: directly soldered on GSB-PCB
GSB - Generic Sensor Board Advanced options on request
• Configurable digital I/Os, e.g.:
Open collector outputs to switch external loads
▪ based on programmable O2-concentration thresholds
▪ programmable switching hysteresis
customized sensor status indication, e.g. error indicator
• Configuration via RS232
flexible configuration of sensor operation parameters
configuration interface for Windows-OS
• Customized GSBs
to implement customer specific digital or analog interfaces
customer specific terminals as replacement for screwed terminals
cost optimization for higher volumes
• Customized firmware
GSB - Generic Sensor Board Basic Specification
Absolute maximum ratings
Values beyond these limits might cause permanent damage
Symbol | Parameter | min. | max | Unit |
Vcc | External power supply voltage | 6 | 25 | V |
Top | GSB operating temperature range | 0 | 50 | °C |
RIout | Load on Iout | 0 | 270 | Ω |
IVout | load current on Vout | 0 | 1 | mA |
Ioc | Open collector sink current (each) | 0 | 50 | mA |
Voc | Open collector voltage | 5 | 30 | V |
VinR232 | Input voltage range R232 | -25 | 25 | V |
Recommended operating conditions
For optimal measurement performance the GSB should be operated at recommended conditions
Symbol | Parameter | min. | max | Unit |
Vcc | External power supply voltage | 11 | 12 | V |
Top | GSB operating temperature range | 15 | 25 | °C |
RIout | Load on Iout | 0 |
|
Ω |
IVout | load current on Vout | 0 |
|
mA |
Iocsum | Open collector sink current (sum) | 1 |
|
mA |
▪ Top refers only to the GSB ambient temperatures, an external connected sensor is not affected by this limitation
▪ Iout refers to the 4-20mA analog output and Vout refers to the 0-5V analog output
▪ Ioc and Voc are only of relevance, if the open collector outputs are used
Analog output levels
Symbol | Output level | Condition | min. | typ. | max | Unit |
Vout | Full scale | recommended conditions, calibrated at full scale | 4.975 | 5 | 5.025 | V |
|
Zero point | 0 | 0.01 | 0.025 | V | |
Iout | Full scale | 19.95 | 20 | 20.05 | mA | |
|
Zero point | 3.95 | 4 | 4.05 | mA |
The basic GSB configuration contains the following preinstalled sensor settings, to be selected via DIP switch. SP-zz-xxx configurations on request
|
|
DIP Switch Bit |
sensor description |
GSB full scale |
O2 calibration concentration |
sensor
voltage
|
heater
control
|
||
sensor type | 1 | 2 | 3 | ||||||
0 | reserved | ON | ON | ON | - | - | - | - |
|
1 | SO-zz-001 | OFF | ON | ON | ppm | 1000 vol.-ppm O2 | 1000 vol.-ppm O2 | 0,70 Volt | Rconst |
2 | SO-zz-010 | ON | OFF | ON | 1% | 1 vol.-% O2 | 1 vol.-% O2 | 0,75 Volt | Rconst |
3 | SO-zz-020 | OFF | OFF | ON | 2% | 2 vol.-% O2 | 2 vol.-% O2 | 0,75 Volt | Rconst |
4 | SO-zz-050 | ON | ON | OFF | 5% | 5 vol.-% O2 | 5 vol.-% O2 | 0,80 Volt | Rconst |
5 | SO-zz-250 | OFF | ON | OFF | 25% | 25 vol.-% O2 | 209 vol.-% O2 | 0,85 Volt | Rconst |
6 | SO-zz-960 | ON | OFF | OFF | 96% | 100 vol.-% O2 | 209 vol.-% O2 | 1,60 Volt | Rconst |
7 | reserved | OFF | OFF | OFF | - | - | - | - |
|
Measurement accuracy – sensor characteristic
If operated at recommended conditions the measurement accuracy will be primarily limited by the sensor accuracy. Also the dynamic sensor characteristic will not be significantly changed by the GSB. See Sensor data sheet for more details.
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Short Probe Zirconia Oxygen Sensors (Zirconia O2 Sensor)
-
Oxygen Analyser
-
Zirconia Oxygen Analyser
-
Zirconia Oxygen Sensor System
-
Screwfit Zirconia Oxygen Sensor
-
Screwfit Zirconia Oxygen Sensor
-
Flange Mount Zirconia Oxygen Sensor
-
Miniature Zirconia Oxygen Sensors
-
Oxygen Sensor Interface Board
-
Zirconia Oxygen Sensor System