Free Chlorine Sensors - FCL502, FCL505 and FCL510

  • Free Chlorine Sensors
  • Free Chlorine Sensors
  • Free Chlorine Sensors
  • Free Chlorine Sensors
Mouse over image to zoom

• Measuring Range: 0-2,0-5 or 0-10 ppm free chlorine
• pH Range: 5.5- 8.5
• Body Material: PVC
• Membrane Cap Material: PVC housing,Proprietary Membrane
• Temperature Range: 0-45 Degrees C
• Maximum Pressure: 14.7 psig (1 Bar)
• Output: 4-20mA

  • Model Number: FCL502, FCL505 and FCL510
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

FCL502, FCL505 and FCL510 Free Chlorine Sensors

Parts covered by this product data sheet include:
FCL502, FCL505, FCL510, FCLA-5015 thru FCLA-5018 FC72, FM001

FCL Series Free Chlorine Sensors Features
• Amperometric Sensor Technology
• Real Time Continuous Measurement
• Ranges: 0-2, 0-5 or 0-10ppm
• Built-In Temperature Compensation
• Large Electrolyte Reservoir
• Replaceable Membrane and Electrolyte

FCL Series Free Chlorine Sensors Description
The FCL series (free chlorine sensors) feature amperometric measurement technology. Designed for disinfection applications in water treatment, for, pools, etc., these sensors are available in three ranges for detecting ppm level of free chlorine. Choose the FCL502 for 0-2ppm, FCL505 for 0-5ppm and FCL510 for 0-10ppm of free chlorine.


Free Chlorine Sensors Typical applications include:

    Industrial wastewater

    Cooling water
    Bottle rinsing
    Irrigation water in agriculture

Sensors output a 4-20mA signal scaled to 0-2, 0-5 or 0-10ppm. Membrane cap and fill solution are easily replaceable to maximize sensor life. 

Sensors can be used in new installations with Sensorex flow cell or installed as replacement for other 4-20mA output  free chlorine sensors. (Consult Sensorex for compatibility).

An optional flow meter (FM001) is available to help provide consistent flow for added system stability.

FCL Series Free Chlorine Sensors Specifications

Measuring Range

0-2,0-5 or 0-10 ppm free chlorine

pH Range

5.5- 8.5

Body Material

PVC

Membrane Cap Material

PVC housing,Proprietary Membrane

Temperature Range

0-45 Degrees C

Maximum Pressure

14.7 psig (1 Bar)

Temp Comp

Integrated

Cable Length

10 feet 2 conductors,tinned leads

Process Connection

flow cell 1/4 inch FNPT threads

Output

4-20mA

Flow Range

Min 0.2 gpm (45 l/hr),Max 0.6 gpm (135 l/hr)

Supply Voltage

12-24V DC,250mA minimum

Cross Sensitivity

Chlorine dioxide,Ozone,bromine,iodine


FCL Series Free Chlorine Sensors Ordering Information

Sensors
FCL502
0-2 ppm Free chlorine 4-20mA sensor, 2 replacement cap & band, 2 each 30mL refill, 
3 pieces cathode polishing squares.

FCLD505
0-5 ppm Free chlorine 4-20mA sensor, 2 replacement cap & band, 2 each 30mL refill, 
3 pieces cathode polishing squares.

FCLD510
0-10 ppm Free chlorine 4-20mA sensor, 2 replacement cap & band, 2 each 30mL refill, 
3 pieces cathode polishing squares.


Accessories
FC72
Flow Cell, 1/4" FNPT inlet and outlet, includes 2 each 1/4" NPT barbed tube fittings 
for 3/8" ID tubing.

FCLA-5015
FCL fill solution, 30mL, 2 each
FCLA-5016
FCL 5 Series sensor replacement membrane cap, 2 each, pressure relief band
FCLA-5017
3 each cathode polishing squares
FCLA-5018
Maintenance kit for FCL5 series sensors includes: 1 replacement cap, 2 each 30mL refill, 
3 each cathode polishing squares & 1 each pressure relief band

FM001
Flow meter, 0.1-1.0 gpm (0.5-4.0 Lpm) range, 1/4" FNPT & 1/2" MNPT threads plus 
(2x) 3/8" tube fittings


If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service