Disposable Intra Uterine Pressure Sensor - MEAS 1630

  • Disposable Intra Uterine Pressure Sensor
  • Disposable Intra Uterine Pressure Sensor
Mouse over image to zoom

• ​Operating Pressure Range: -50 ~ 300mmHg
• ​Sensitivity: 4.9~5.1uV/V/mmHg
• ​Warm-Up Time: 5 Seconds
• ​Frequency Response​: 1200Hz
• ​1% Accuracy for Replacements
• ​5.0 uV/V/mmHg Sensitivity
• ​Operating Temperature10°C to 40°C
• ​Humidity (External)​: 10-90% (non-condensing)​

  • Model Number: MEAS 1630
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

1630 Medical Pressure Sensor Specifications
• Disposable Intra Uterine Pressure Sensor
• Disposable Blood Pressure Sensor
• AAMI Specifications
• Low Cost Disposable
• Dielectric Gel Barrier
• Fully Tested & Calibrated

The 1630 is a fully piezoresistive silicon pressure sensor for use in intrauterine pressure monitoring. The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer's disposable intrauterine housing. The sensor is designed to meet the requirements as described in the Association for the Advancement of Medical Instrumentation (AAMI) specification for Blood Pressure Transducers.

The pressure sensor consists of a pressure sensing element mounted on a ceramic substrate. Thick film resistors on the ceramic substrate are laser trimmed for compensation and calibration. A plastic cap is attached to the ceramic substrate to provide an easy method of attachment to the customer’s assembly and protection for the sensing element. A dielectric gel is placed over the sensor to provide fluid isolation.

The 1630 pressure sensors are batch manufactured in a 3x8 element array on a ceramic substrate (24 units per substrate). The products are shipped in anti-static shipping containers. They can also be shipped on a tape and reel. Performance characteristics and packaging can be easily tailored on a special order basis to meet the requirements of specific customers.

1630 Medical Pressure Sensor Features
• Low Cost, Medical Applications
• Small Size and Reliable Performance
• Gel Isolation for Liquids
• Operates from 10°C to 40°C
• Compatible for Automated Assembly
• 1% Accuracy for Replacements
• 5.0 uV/V/mmHg Sensitivity
• Customization for OEM Applications

1630 Medical Pressure Sensor Applications
• Intrauterine Monitoring
• Intensive Care Units
• Infusion Pumps
• Kidney Dialysis Machines
• Vacuum Assisted Birth
• Surgical Procedures

1630 Medical Pressure Sensor Performance Specifications
Supply Voltage: 5.0 Vdc
Ambient Temperature: 23°C (unless otherwise specified)

Parameters
Min
Typ
Max
Units Notes
Operating Pressure Range
-50
300 mmHg

Over Pressure
-400
1200 mmHg

Zero Pressure Offset 
-20
20 mmHg

Sensitivity
4.9 5.0 5.1 uV/V/mmHg

Linearity and Hysteresis (0 to 300 mmHg)
-2.0
2.0 %Span
1
Input Impedance
1200
3200

Output Impedance
270
330

Supply Voltage
2 5 10 Vdc or Vac rms

Leakage Current (@ 120 Vac rms, 60Hz)


2 uA

Warm-Up Time

5
Seconds

Frequency Response

1200
Hz

Offset Drift


2 mmHg 
2
Thermal Span Shift
-0.1
0.1 %/°C
3
Thermal Offset Shift
-0.3
0.3 mmHg/°C
3
Phase Shift (@ 5KHz)


5 Degrees

Sterilization (ETO)
3

Cycles
4
Operating Temperature
10
40 °C

Storage Temperature
-25
+70 °C

Operating Product Life


72 Hours

Shelf Life
3

Years

Humidity (External)
10-90% (non-condensing)

Media Interface
Dielectric Gel

Notes
1.  Best fit straight line.
2.  Over an 8 hour time period after a 10 minute warm-up.
3.  Over operating temperature range 10-40°C with respect to 23°C.
4.  Sterilization performed by customer.