STs Series pH Sensor - 1014 pH

  • STs Series pH Sensor
Mouse over image to zoom

• Combination double-junction electrode measuring 0 to 14 pH
• Large reference half-cell to reduce calibration frequency
• Secondary 0-50°C temperature sensor
• 2-wire, 4-20 mA analog outputs compatible with wide range of dataloggers or remote telemetry units (RTU)
• Narrow diameter (0.75 inches or 1.9 cm) for groundwater monitoring applications
• Low power usage

  • Model Number: 1014 pH
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

Description:

    The pHionics STs SeriespH is a dual-output sensor that pairs with a datalogger or RTU for real-time water quality monitoring.  A narrow diameter allows the STs Series to go where other sensors cannot while a durable, chemical-resistant housing ensures many years in-use.  

    All parts, including the cable, are easily replaceable without tools usingpHionicspatentedpHiConnSystem, saving frustration and money.

    In addition, isolation, differential amplification, and shielding ensure a strong signal with high signal-noise ratios, all while remaining low-power.

    These features make our pH sensor well-suited for applications such as groundwater monitoring, wastewater treatment, aquaculture, pollution monitoring, etc.

 

Highlighted Features:

Combination double-junction electrode measuring 0 to 14 pH

Large reference half-cell to reduce calibration frequency

Secondary 0-50°C temperature sensor

2-wire, 4-20 mA analog outputs compatible with wide range of dataloggers or remote telemetry units (RTU)

Narrow diameter (0.75 inches or 1.9 cm) for groundwater monitoring applications

Low power usage

 

Applications:

Groundwater Monitoring

    o Agricultural Runoff

    o Mining Pollution

Industrial Effluent Monitoring

Wastewater Treatment

Ballast Monitoring

Environmental Remediation

Aquaculture

 

Specifications:

pH Sensor

Output

4-20 mA

pH Range

0-14 pH

Operational Temperature Range

0-70

Electrode Type

Silver/silver chloride, double junction, replaceable gel electrode

Transmitter Type

2-wire

Pressure Range

0-70 psi

Automatic Temperature Compensation

0-70

Isolation

600 VDC

Power Supply Voltage

8-40 VDC

Warm-up Time

2 seconds

Response time

95% < 5 seconds

Wetted Materials

316 Stainless Steel, Delrin®, Viton, glass, epoxy

Length

14.25 in (36.20 cm)

Width

0.750 in (1.9 cm) maximum

Weight (No cable)

0.50 lbs (0.227 kg)

Linearity

±0.014 pH*

Accuracy

±0.014 pH*

Sensitivity

±0.01 pH

Stability

±0.03 pH per year

Temperature Sensor

Output

4-20 mA

Temperature Range

0-50

Accuracy

+/- 1

Linearity

+/- 0.5

Power Supply Voltage

7-40 VDC

Cable

Cable Material

Polyurethane

Type

4 conductor, Kevlar® reinforced, water blocked, shielded, twisted-pair wiring

Maximum Distance

3 miles

Resistance per 1000 ft (304.8 m)

26 Ohms

Replaceable?

Yes

*After proper calibration


If you cannot find what you want, you can entrust OFweek to source for you. Just click: