Indoor Air Quality Monitors - 9301P-OEM

  • Indoor Air Quality Monitors
Mouse over image to zoom

• ​9301P-OEM: Measures 0.3 µm to 25 µm
• ​0.1 CFM (2.83 LPM) flow rate
• ​Long life laser diode technology
• ​Measures up to 30 channels of simultaneous data

  • Model Number: 9301P-OEM
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

    The 9301P-OEM measures up to 30 channels of particle count data from 0.3 to 25.0 µm with a flow rate of 0.1 CFM (2.83 LPM).The assembly is mounted on a plate for ease of integration. The unit includes a particle sensor, particle counter board, integrated vacuum pump and manifold with pre/post filters, real-time clock/calendar with battery backup, logging and data storage for up to 45,000 sample records, communicates via UART and operates from 7-15 vdc.

    All Particles Plus counters meet ISO 21501-4 and JIS B9921. The 9301P-OEM ensures compliance with an on-board pulse height analyzer.

 

Features:

• 9301P-OEM: Measures 0.3 µm to 25 µm

• 0.1 CFM (2.83 LPM) flow rate

• Long life laser diode technology

• Measures up to 30 channels of simultaneous data

• Particle concentrations up to I 0,000,000 Particles/ft3 @ I 0% coincidence loss

• Internal vacuum pump with manifold and pre/post filters

• Stores up to 45,000 sample records for on-board data redundancy

• User-selectable channel sizes

• Complies with ISO 21501-4 and JIS B9921 standards

• Assembly mounted on back plate for ease of integration

• I year warranty. Extended warranties available.

 

Specifications:

Model

9301P-OEM

Size Range

0.3 to 25µm

Size Channels

Factory calibrated at 0.3,0.5,1.0,2.5,5.0,10.0µm

Flow Rate/Accuracy

0.1CFM(2.83 LPM)/+/-3%

Aerosol Concentration Range

0.01 to 20000 ug/m3

Light Source

Long life laser diode

Resolution

.00000003 mg/m3

ZeroStability

No appreciable drift

Calibratio

NIST traceable

Internal Vacuum Pump

Intemal pump with automatic flow control

Number of Channels

30

Custom Size Channels

Calibration for custom size channels available

Airflow

Internally monitored

Audible Alarm

Adjustable built-in alarm

Communication Modes

Universal asynchronous receiver-transmitter (UART)

Standards

ISO 21501-4 and JIS B9921

Instrument Calibration

Recommended minimum once per year

Dimensions (LxWx H)

3.41"x4.89"x1.57(8.7 cmx12.4 cmx4.0 cm)includes inlet height

Weight

13oz.

Data Storage

65,000 sample records(rotating buffer including particle count data and environmental data

Sample Time

1 second to 99 hours

Power

7-15 VDC

Operating Conditions

41°to 104F(5°to 40C)/20% to 95% non-condensing

Storage Conditions

32°to 122F (0°to 50C)/Up to 98% non-condensing

Warranty

1 year limited warranty. Extended warranties available.

 

If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service