Hand-held Temperature and Humidity Measuring Device - XP200/XP201

  • Hand-held Temperature and Humidity Measuring Device
Mouse over image to zoom

Measurement range: - 40 … 80 ° C
Accuracy (20°C): ± 0.1°C
Accuracy (0...40°C): ± 0.2°C otherwise ± 0.5°C
Resolution: 0.01°C
Principle: PT1000, Class A, DIN EN 60751

  • Model Number: XP200/XP201
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

Hand-held measuring device XP200/XP201 for measuring temperature and humidity


X-pert range for humidity and temperature measurements in climate and environmental technology.

- TFT LCD, anti-glare colour display
- Capacitive touch screen
- Sampling rate 1s
- Data recording simultaneously on up to 3 channels
- Graphical analysis including standard deviation
- Integrated  ash memory with space for up to 200 data blocks, or up to three hours continuous recording
- USB port for data transfer to Smartgraph 3 (included in delivery)
- Multilingual interface
- Numerous dimensions calculated
- Online  rmware update


Hand-held measuring device XP200

Technical Data Dimensions Length 85 mm, Ø 12 mm
Weight Approx. 50g
Protection Polycarbonate / IP65
Permitted operation temp. 0...50°C
Permitted humidity 0...100% RH
Storage temperature -20...60°C
Storage humidity 20...80% RH
Relative Humidity Measurement range 0… 100% RH
Accuracy ± 2 % (0 … 90 %), ± 3 % (90 … 100 %) RH
Resolution 0.01% RH
Principle Capacitive
Temperature Measurement range - 40 … 80 ° C
Accuracy (20°C) ± 0.1°C
Accuracy (0...40°C) ± 0.2°C otherwise ± 0.5°C
Resolution 0.01°C
Principle PT1000, Class A, DIN EN 60751
Absolute Humidity Measurement range 0...300g/m3
Unit g/m3
Dew Point Temp. Measurement range -40...80°C
Mixing Ratio Measurement range 0...550g/kg
Compatibility XA1000, XP200, OPUS20E
Accessories Stainless steel sinter cap 5120.212
Calibration salt 11,3% RH 5700.113
Calibration salt 32,8% RH 5700.328
Calibration salt 52,9% RH 5700.529
Calibration salt 75,3% RH 5700.753
Calibration salt 90,1% RH 5700.901
Calibration adapter 8120.ADAP
Technical data  Dimensions 170 x 62 x 34 mm
Weight Approx. 205g
Storage conditions Permitted ambient temperature -20...60°C
Permitted rel. humidity <90% RH non-condensing
Operating conditions Permitted rel. humidity <90% RH (20g/m3) non-condensing
Permitted altitude above sea level 4000m
Power supply Power supply 4 Alkaline LR6 AA 1.5V / USB 5V
Active power consumption Approx. 400mW
Battery life passive Approx. 1 year
Battery life active Min. 24 hours
Sensor power supply 5.5V ± 10% DC, max. 200mA
Data storage Integrated data storage Up to 200 data/approx. 1 Mio measured values
Interface USB Cable and SmartGraph3 software included
Resolution De nition of measured values 2 decimal places
Display Control Touch screen, capacitive
Surface, toughened glass Degree of hardness: 7, scratch-resistant
Integrated air pressure sensor Measuring range (full accuracy) 800...1,100mbar
Accuracy at 25°C,1013.25mbar 0.5mbar
Long-term stability typ. - 1mbar/year
Measurement resolution 0.024mbar
Measuring principle Piezoresistive
Calculated measurement categories Mathematical: MIN/MAX/AVG/HOLD
Temperature (°C/°F)
Rel. humidity (% RH)
Rel. humidity of ice (% RH)
Water vapour density (absolute humidity) g/m3
Dew point temperature °C/°F
Frost point temperature °C/°F
Mixing ratio at saturation (100%) g/kg
Volume fraction of water vapour /mass fraction of water vapour (%)
Wet-bulb temperature °C/°F
Ice-bulb temperature °C/°F
Speci c Enthalpy (mass of air) kJ/kg
Saturation vapour pressure above ice/water (hPa)
Water vapour particle pressure (hPa)
Air density kg/m3
Accessories Extension and/or connecting cable for digital sensor, 2m
Extension and/or connecting cable for digital sensor, 10m


If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service