• Measurement Range: 0 to 50 ppm
• Resolution: < 5 ppb (instrumentation dependent)
• Repeatability: < +/- 3 % of reading
• Response Time: < 15 seconds typical
• Operating Temperature: -30 to 55°C
• Operating Humidity: 0 to 100% RH non-condensing
• Power Consumption: 10 to 50 uW

  • Model Number: 3SP_H2S_50 P
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

3SP_H2S_50 P
15x15 H2S Sensor 50 ppm Package 110-303


3SP_H2S_50 P Hydrogen Sulfide Sensor H2S Sensor Description
SPEC's printed sensors offer the performance of the best quality electrochemical sensors at a fraction of the price. SPEC's printed sensors are also ultra-thin, offering easy integration into wireless, portable, and networked solutions. These sensors are ideal for health, environmental, industrial and residential monitoring, because of their high performance, low cost and small size.


3SP_H2S_50 P Hydrogen Sulfide Sensor H2S Sensor Benefits

• Small Size with Low Profile (20x20 x3.0mm)
• Long Life (10 years expected life)
• Fast Response (< 15 seconds)
• Low Power Consumption
• Individually Calibrated (NIST Traceable)
• ROHS Compliant


3SP_H2S_50 P Hydrogen Sulfide Sensor H2S Sensor Applications

• Fixed Industrial Safety Monitoring
• Portable Industrial Safety Detection
• Portable Personal Safety Monitor
• Indoor Air Quality Monitoring
• Outdoor Air Quality Monitoring
• Bad Breath/Halitosis Respiration Sensing


3SP_H2S_50 P Hydrogen Sulfide Sensor H2S Sensor Specifications

Measurement Range

 0 to 50 ppm

Resolution

 < 5 ppb (instrumentation dependent)

Repeatability

 < +/- 3 % of reading

Response Time

 < 15 seconds typical

Sensitivity @ 0 mV bias

 212 +/- 87 nA/ppm

Expected Operating Life

 > 5 years (10 years @ 23+/-3C; 40+/-10% RH)

Operating Temperature Range

 -30 to 55°C (-20 to 40°C recommended)

Operating Humidity Range - non-condensing

 0 to 100% RH (15 to 95% recommended)

Power Consumption

 10 to 50 uW (circuit & ambient H2S dependent)


3SP_H2S_50 P Hydrogen Sulfide Sensor H2S Sensor Cross Sensitivity

Gas/Vapor

Concentration

Typical Response PPM H2S

Methane

500 ppm

0.1

Ammonia

100 ppm

0.1

Nitrogen Dioxide

10 ppm

-2.0

Hydrogen Sulfide

25 ppm

25

Carbon Monoxide

400 ppm

1.1

Ozone

5 ppm

-0.9

Sulfur Dioxide

20 ppm

1.7

Nitric Oxide (NO)

50 ppm

1.2

Chlorine

10 ppm

-2.2

n-Heptane

500 ppm

< 0.05