Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor - OKS-11
• Measures Oxygen From 0 to 100%
• Electrochemical Fuel Cell Sensor Technology
• Self-Contained, No Need To Add Electrolyte
• Excellent Stability, Performance, & Linearity
• Competitive Drop-In Field Replacement
- Quantity:
- - +
Product Specification
1.OKS-11 Electrochemical Oxygen Sensor Descriprion:
• Measures Oxygen From 0 to 100%
• Electrochemical Fuel Cell Sensor Technology
• Self-Contained, No Need To Add Electrolyte
• Excellent Stability, Performance, & Linearity
• Competitive Drop-In Field Replacement
The OKS-11 oxygen sensor provides high levels of accuracy, reliability, and linearity throughout its measurement ranges. The oxygen sensor is based on the galvanic electrochemical fuel cell principle and is manufactured in-house under a strict quality procedure.
Backed with excellent customer service and an industry proven product design, our sensors excel in performance and quality while reducing the cost of ownership.
Varying performance advantages to the Southland Sensing sensors can be attributed to minor modifications in dimensions, connections, and packaging. However, the principle of operation and the basic components remain the same.
Along with excellent performance and exceptional quality, Southland Sensing offers rapid delivery times to a global customer base for most oxygen sensors.
2.OKS-11 Electrochemical Oxygen Sensor Specifications:
Sensor Technology |
Electrochemical |
Signal Output1 |
13 - 16 mV |
Measuring Range |
0 - 100 Percent |
Response Time T90 |
13 Seconds |
Accuracy Full Scale2 |
± 1% of Signal |
Repeatability F.S. |
± 0.5% |
Temperature |
Compensated |
Operating Temp |
0 to 45 ℃ |
Storage Temp |
0 to 45 ℃ |
Cross Sensitivity |
ISO 80601-2-55 |
Housing Material |
White ABS |
Humidity |
0 - 100% RH(Non - Condensing) |
Expected Life3 |
< 60 Months |
Storage |
12 Months |
Warranty4 |
12 Months |
1. Signal Output is measured in air at 25 º C, Sea Level.
2. Full Scale Accuacy is calculated with constant pressure, temperature and proper calibration (80% O2 Value on
Full Scale Range). Drastic Temperature change can result in a maximum error of ± 10%.
3. Expected life is calculated when O2 < 20.9% @ 25 º C, Sea Level.
4. Southland Sensing Ltd. Warrants the sensors for the period noted above to be free from defects in materials and workmanship. Oksidyne will not be held liable for sensors damaged due to customer neglect.
5. Southland Sensing Ltd. sensors are tested and validated to BSI Standard BS EN 50104:2010
3.OKS-11 Electrochemical Oxygen Sensor Electrical & Gas Connection:
M16 x 1.0mm Thread w/ O’ring
3-Pin Molex Header (MOLEX 22-29-2031)
4.OKS-11 Electrochemical Oxygen Sensor Electrical & Gas Intended Use:
These oxygen sensors are designed to be used to monitor the partial pressure of oxygen in breathing gas mixtures for anaesthesia, ventilators, medical oxygen concentrators, incubators, and general oxygen monitors.
5.OKS-11 Electrochemical Oxygen Sensor Electrical & Gas Dimensions:
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Oxygen Analyzer
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Oxygen(O2)Gas Sensor
-
Oxygen(O2)Gas Sensor
-
High Quality O2 Sensors