LuminOx Sealed Optical Oxygen Sensor (O2 Sensor) - LOX-02-S

  • LuminOx Sealed Optical Oxygen Sensor (O2 Sensor)
  • LuminOx Sealed Optical Oxygen Sensor (O2 Sensor)
Mouse over image to zoom

• Supply voltage: 4.5 - 5.5V DC
• Supply current: <7.5mA (streaming one sample per second), <20mA Peak
• Output Type: 3.3V TTL level USART
• Operating Temperature: -30°C to +60°C
• Humidity​: 0 - 99% Rh (non-condensing)​
• Barometric pressure range 500 to 1200mbar

  • Model Number: LOX-02-S
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

LOX-02-S LuminOx Sealed Optical Oxygen Sensor General Description
The LuminOx Sealed Optical Oxygen Sensor Range Offers Users With Application Requirements To Only Measure Sample Gas At The Front Face Of The Sensor.

The LuminOx Sealed Optical Oxygen Sensor is developed and manufactured by SST, it is part of the LuminOx family, which is a range of factory calibrated oxygen sensors which measure ambient oxygen levels using the principle of luminescence quenching.

The LuminOx Sealed Optical Oxygen Sensor (LOX-02-S) is an extension to the standard range of LuminOx oxygen sensors. It is a similar diffusion type sensor like the original LOX-02 sensor. However, the LOX-02-S has a sealed base that is gas-tight, meaning only the face of the sensor with the protective membrane needs to be in the sample gas environment, externally leaving easy access to the pins of the sensor.

The patented technology allows SST to produce a range of sensors that provides low power operation, while offering a much longer lifetime than electrochemical and galvanic sensors.

The sensors each have an internal sensor for temperature and barometric pressure compensation, meaning the customer does not need to consider this in their design. It also communicates directly to the customers USART so no signal conditioning circuitry is required.

LuminOx is a completely solid state sensor and contains no lead or liquid electrolytes. It is therefore 100% RoHS compliant. It also has virtually no cross sensitivity, allowing it to be used in applications where other technologies are unreliable due to the particular mix of gases present.

The luminescence quenching principle measures the partial pressure of oxygen (ppO2). The internal barometric pressure sensor measurement along with the ppO2 measurement allows an oxygen concentration (O2%) value to be calculated. The USART output allows the sensor to communicate the values of O2%, ppO2, barometric pressure, internal temperature and a sensor status.

LOX-02-S LuminOx Sealed Optical Oxygen SensoFeatures
• Luminescence-based optical technology, NOT electrochemical
• Contains no hazardous materials; RoHS & REACH compliant
• Connects directly to a microcontroller without any additional circuitry
• Factory calibrated
• High accuracy
• Maintenance free

LOX-02-S LuminOx Sealed Optical Oxygen SensoBenefits
• Compact, diffusion-type housing with sealed base
• Low power, long life due to non-depleting sensing principle
• Low cost

LOX-02-S LuminOx Sealed Optical Oxygen SensoTechnical Specifications

Supply voltage (Vs)
4.5-5.5V DC
Supply current (ls)
<7.5mA (streaming one sample per second), <20mA Peak
Output Type
3.3V TTL level USART
Temperature

    Operating: -30°C to +60°C
    Storage: -30°C to +60°C
Humidity
0 - 99% Rh (non-condensing)
Barometric pressure range
500 - 1200mbar

LOX-02-S LuminOx Sealed Optical Oxygen SensoOutput Values
Oxygen range
0 - 25% O2
Oxygen pressure range
0 - 300mbar ppO2
Response time
T90 < 30s (typical)
Accuracy

    ppO2 < 2% FS
    Temperature Indication only
    Pressure ±5mbar
    O2 Determined by ppO2 & pressure accuracy
Resolution

    ppO2 0.1mbar
   Temperature
0.1°C
    Pressure
1mbar
   O2
0.01%
Lifetime
> 5 years

If you cannot find what you want, you can entrust OFweek to source for you. Just click:

Sourcing Service