Chlorine Sensor (CL2 Sensor)
-
Price
Price on request ≥1piece
-
Model Number
CL2-A1 -
Data Sheet
-
Package
Original packaging -
Specs
CL2-A1 -
Quantity
- +Send Inquiry
Find Similar Products
Product Description
CL2-A1 Chlorine Sensor (CL2 Sensor) Description
Alphasense Chlorine gas sensors operate using proven fuel cell technology.
CL2-A1 Chlorine Sensor size:
20mm diameter, the industry standard size for portable gas detectors
The Cl2 range provides OEMs with reliable sensors for use in a number of high volume applications. Strong signal levels combined with low zero current allows resolution to 20ppb and an operating range from +0 to 20ppm. The sensors are suitable for inclusion in fixed installation sensing heads and portable safety instruments.
These sensors have improved transient response to humidity changes compared to competitors' equivalent sensors.
The Cl2 sensors offer our electrolyte leak-free guarantee and reliable long-term detection performance. The leakproof housing is moulded with a colour coded top for ease of identification.
Matched Transmitter board for CL2-A1 CL2 Sensor

4-20mA Digital Transmitter Board
• Factory calibration, custom-set for immediate use
• Digital zero, sensitivity and temperature compensation
• Digital and 4-20mA output
• Biased or unbiased operation
• Calibration and range change
Alphasense Chlorine gas sensors operate using proven fuel cell technology.
CL2-A1 Chlorine Sensor size:
20mm diameter, the industry standard size for portable gas detectors
The Cl2 range provides OEMs with reliable sensors for use in a number of high volume applications. Strong signal levels combined with low zero current allows resolution to 20ppb and an operating range from +0 to 20ppm. The sensors are suitable for inclusion in fixed installation sensing heads and portable safety instruments.
These sensors have improved transient response to humidity changes compared to competitors' equivalent sensors.
The Cl2 sensors offer our electrolyte leak-free guarantee and reliable long-term detection performance. The leakproof housing is moulded with a colour coded top for ease of identification.
CL2-A1 Chlorine Sensor (CL2 Sensor) PERFORMANCE |
||
Sensitivity |
nA/ppm in 10ppm Cl2 |
-350 to -750 |
Response time |
t90 (s) from zero to 10ppm Cl2 (33Ω load resistor) |
< 60 |
Zero current |
ppm equivalent in zero air |
± 0.3 |
Resolution |
RMS noise (ppm equivalent, 33Ω load resistor) |
< 0.02 |
Range |
ppm limit of performance warranty |
20 |
Linearity |
ppm error at full scale, linear at zero and 5ppm Cl2 |
± 1.5 |
Overgas limit |
maximum ppm for stable response to gas pulse |
50 |
|
|
|
CL2-A1 Chlorine Sensor (CL2 Sensor) LIFETIME |
|
|
Zero drift |
ppm equivalent change/year in lab air, monthly test |
< 0.05 |
Sensitivity drift |
% change/year in lab air, monthly test |
< 10 |
Operating life |
months until 80% original signal (24 month warranted) |
> 24 |
|
|
|
CL2-A1 Chlorine Sensor (CL2 Sensor) ENVIRONMENTAL |
||
Sensitivity @ -20°C |
% (output @ -20°C/output @ 20°C) @ 10ppm Cl2 |
65 to 85 |
Sensitivity @ 50°C |
% (output @ 50°C/output @ 20°C) @ 10ppm Cl2 |
105 to 125 |
Zero @ -20°C |
ppm equivalent change from 20°C |
< ± 0.2 |
Zero @ 50°C |
ppm equivalent change from 20°C |
< 0 to -0.8 |
|
|
|
CL2-A1 Chlorine Sensor (CL2 Sensor) CROSS SENSITIVITY |
||
H2S sensitivity |
% measured gas @ 20ppm H2S |
< -300 |
NO2 sensitivity |
% measured gas @ 10ppm NO2 |
100 |
NO sensitivity |
% measured gas @ 50ppm NO |
< 3 |
SO2 sensitivity |
% measured gas @ 20ppm SO2 |
< -8 |
CO sensitivity |
% measured gas @ 400ppm CO |
< 0.1 |
H2 sensitivity |
% measured gas @ 400ppm H2 |
< 0.1 |
C2H4 sensitivity |
% measured gas @ 400ppm C2H4 |
< 0.1 |
|
|
|
CL2-A1 Chlorine Sensor (CL2 Sensor) KEY SPECIFICATIONS |
||
Temperature range |
°C |
-20 to 50 |
Pressure range |
kPa |
80 to 120 |
Humidity range |
%rh continuous |
15 to 90 |
Storage period |
months @ 3 to 20°C (stored in sealed pot) |
6 |
Load resistor |
Ω (for optimum performance) |
33 |
Weight |
g |
< 6 |
Matched Transmitter board for CL2-A1 CL2 Sensor

4-20mA Digital Transmitter Board
• Factory calibration, custom-set for immediate use
• Digital zero, sensitivity and temperature compensation
• Digital and 4-20mA output
• Biased or unbiased operation
• Calibration and range change