Product Description
RADIKON
Versatile radiation controller for industry and science
RADIKON Versatile Radiation Controller Properties
The RADIKON is used to control the value generated by a detector (e.g. radiation sensor or a pressure sensor). If the detector value falls below a certain setpoint value a relay is activated where a valve, an alarm buzzer or other modules can be connected.
The highlight of the RADIKON is its high versatility making the module very interesting to industrial developers who need to match the controller with different imput signals.
The RADIKON can read the generic current of a photodiode (5nA < ISensor < 5µA) as well as 0...x Volt and 0(4)...20mA probe signals. The value x is user definable. The RADIKON outputs a voltage of 0…10V which can be used to attach a separate display or to connect the RADIKON with a PID controller. A multi-colour LED changes its colour with the radiation intensity level. All these features as well as easy set up and configuration make the RADIKON a perfect tool for developers or manufacturers of small series. For medium series controller modules with customized properties are offered.
RADIKON Versatile Radiation Controller Specifications
Versatile radiation controller for industry and science
RADIKON Versatile Radiation Controller Properties
The RADIKON is used to control the value generated by a detector (e.g. radiation sensor or a pressure sensor). If the detector value falls below a certain setpoint value a relay is activated where a valve, an alarm buzzer or other modules can be connected.
The highlight of the RADIKON is its high versatility making the module very interesting to industrial developers who need to match the controller with different imput signals.
The RADIKON can read the generic current of a photodiode (5nA < ISensor < 5µA) as well as 0...x Volt and 0(4)...20mA probe signals. The value x is user definable. The RADIKON outputs a voltage of 0…10V which can be used to attach a separate display or to connect the RADIKON with a PID controller. A multi-colour LED changes its colour with the radiation intensity level. All these features as well as easy set up and configuration make the RADIKON a perfect tool for developers or manufacturers of small series. For medium series controller modules with customized properties are offered.
RADIKON Versatile Radiation Controller Specifications
Input Values Parameter |
Value |
Generic sensor signal,
e.g. photodiode current
|
0 to x nA |
x can be configured (with jumpers) to 50 nA, 150 nA, 500 nA, 1.5 µA and 5 µA, other values are possible by soldering an own feedback resistor |
|
Voltage input |
0 to x Volt |
x can be configured (with jumpers) to 50 mV, 142 mV, 450 mV, 1160 mV, 2500 mV and 10 V, other values are possible by soldering an own feedback resistor |
|
Current input (current loop) |
y to 20 mA |
y can be configured (with jumpers) to 0 mA or 4 mA |
|
Adjustable setpoints |
relay setpoint, multi colour LED |
|
|
Other Properties |
|
Inputs |
DC-isolated, EMC and EDS conform |
Power supply |
12 to 24 VDC |