Digital Gas Flow Sensor
-
Price
Price on request ≥1piece
-
Model Number
MFM2040U -
Data Sheet
-
Package
Original packaging -
Specs
MFM2040U -
Quantity
- +Send Inquiry
Find Similar Products
-
Gas Flow Sensors FSG4000 Series
Model Number:FSG4000
MOQ:1
-
MEMS Mass Flow Sensor
Model Number:FS4003
MOQ:1
-
Gas Mass Flow Sensor
Model Number:FS4008
MOQ:1
-
MEMS Mass Flow Sensors
Model Number:FS4001
MOQ:1
-
Gas Flow Sensors
Model Number:FS1015CL
MOQ:1
-
MEMS Mass Flow Sensors
Model Number:FS5001L Series
MOQ:1
Product Description
MFM2040U Digital Gas Flow Sensor DESCRIPTION
MFM2000 series is a gas flow sensor with low pressure drop capable of measuring medical gases up to 45 lpm. It is configured to be customizable with different flow inlet and outlet configurations.
MFM2040U Digital Gas Flow Sensor FEATURES
• Low Pressure Drop Flow Sensor
• High Accuracy +/-2.5% Reading
• Calibrated and Temperature Compensated for Air
• Low Power Consumption
• Digital I2C
• Made for Medical Applications
• Calibrated for Air, and O2
MFM2040U Digital Gas Flow Sensor FLOW PERFORMANCE
Measurements performed with air at 23°C, 1 Atm, 50% RH, at 3.3V DC power supply, unless otherwise specified. Calibration conditions for standard liters per minute slm: 20°C, 1013mbar.
ELECTRICAL SPECIFICATIONS (VDD=3.3V, TA=25C, UNLESS OTHERWISE NOTED)
MFM2040U Digital Gas Flow Sensor MECHANICAL SPECIFICATIONS
MFM2000 series is a gas flow sensor with low pressure drop capable of measuring medical gases up to 45 lpm. It is configured to be customizable with different flow inlet and outlet configurations.
MFM2040U Digital Gas Flow Sensor FEATURES
• Low Pressure Drop Flow Sensor
• High Accuracy +/-2.5% Reading
• Calibrated and Temperature Compensated for Air
• Low Power Consumption
• Digital I2C
• Made for Medical Applications
• Calibrated for Air, and O2
MFM2040U Digital Gas Flow Sensor FLOW PERFORMANCE
Measurements performed with air at 23°C, 1 Atm, 50% RH, at 3.3V DC power supply, unless otherwise specified. Calibration conditions for standard liters per minute slm: 20°C, 1013mbar.
Parameter |
Condition |
Min |
Typical |
Max |
Unit |
Flow Range |
Calibration gas: Air |
0.04 |
|
40 |
SLM |
Calibrated Temperature Range |
|
0 |
|
50 |
°C |
Operating Temperature Range |
|
-40 |
10 | 85 |
°C |
Storage Temperature Range |
|
-40 |
|
105 |
°C |
Accuracy |
25ºC +/-2ºC |
|
+/-2.5 |
|
% Reading |
Temperature Error |
0-50ºC |
|
+/-2.5 |
|
reading/°C |
Resolution |
I2C |
|
15 |
|
bit |
Sampling Time |
|
|
7.5 |
|
ms |
Power Consumption |
@ Recommended Heater
Power at zero Flow
|
|
|
7.0 |
mA |
PD |
|
|
650 |
uA |
|
Pressure Drop |
Max flow without mesh |
|
|
60 |
Pa |
Max flow(with mesh screen) |
|
|
200 |
Pa |
|
Operating Pressure Range |
Absolute |
0.7 |
|
1.3 |
Bar |
Operating Overpressure |
Absolute |
|
+/-0.2 |
|
Bar |
Averaging |
10 frames running window averaging |
||||
Wetted Materials |
PES (Medical grade: biocompatible; ISO 10993 or USP Class VI), FR4, Silicon
Nitrite, Silicon, Epoxy, Gold.
|
||||
Standards |
RoHS, REACH, ISO15001, ISO10993-1, ISO10993-5,UL94 V0, IEC |
ELECTRICAL SPECIFICATIONS (VDD=3.3V, TA=25C, UNLESS OTHERWISE NOTED)
Parameter |
Condition |
Min |
Typical |
Max |
Unit |
Supply DC Voltage |
DC |
2.7 | 3.3 | 5.5 | V |
Power Consumption |
@ Recommended Heater Power at
zero Flow
|
|
|
7.0 |
mA |
PD |
|
|
650 |
uA |
|
Data format Scaling Factor |
User read count value from register 0,
then multiply by 1.25 to get actual
flow rate in sccm.
|
1.25 |
sccm |
MFM2040U Digital Gas Flow Sensor MECHANICAL SPECIFICATIONS
Specification |
Condition |
Min |
Typical |
Max |
Unit |
Burst Pressure |
Over ambient |
1 |
|
|
Bar |