PPM O2 Oxygen Sensor Microoxygen sensor - GPR-12-333-H
O-10PPM
Microoxygen analyzer
- Quantity:
- - +
Product Specification
-
This sensor is specifically designed to for Analytical Industies Inc., trace oxygen analyzer, explosion proof, intrinsically safe transmitters and portable analyzers.
-
GPR-1600
GPR-1200
GPR-1100
GPR-18
GPR-1800 Series
GPR-1500 Series
PPM O2 Oxygen Sensor Microoxygen sensor GPR-12-333-H Technical Specifications:
Measuring Range |
0-25% O2 |
Signal Output (1) |
590-950 µA |
Response Time 90% |
3 Seconds |
Accuracy Full Scale (2) |
+ 1% |
Accuracy FS Over Operating Range (3) |
+ 5% |
Drift % Signal Month (2) |
< 1% |
Linearity (2) |
+ 1% |
Humidity Non-Condensing |
0-99% RH |
Temperature Coefficient |
2.54% |
Operating Temperature |
0-45℃ |
Operating Pressure |
Constant 5-30 PSIG |
Flow Rate |
.1-10 LPM |
Expected Life |
12-24 Months O2 <1000 PPM |
Storage (4) |
0-40℃ |
Warranty (5) |
12 Months |
Electrical Connections |
PCB 2 Ring |
Packaging |
Polyetylene bag in 3" x 3" x 1.5" paper box |
(1) In air @ 25℃and 1 atm. Expected life varies inversely with changes in O2 concentration, temperature, pressure.
(2) Under constant conditions. Calibrate as near sample conditions as possible. Calibrate with 95-100% oxygen for measurements > 40% oxygen.
(3) Once the signal output stabilizes following a step change in temperature.
(4) Recommended 0-40℃. Sensors may be stored at -10 to 55℃on an intermittent basis only.
(5) Ex-factory unless agreed to in writing. Under normal application conditions sensor is warranted to be free of defects in material and workmanship for the period specified above proved the sensor is properly installed and operated. Analytical Industries Inc. shall not be liable for buyer's negligence,misapplication, alteration or abuse. The sole remedy for a sensor determined to be defective by Analytical Industries Inc. is limited to replacing the defective sensor.
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Oxygen Analyzer
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Medical Electrochemical Galvanic Fuel Cell Percent Oxygen Sensor
-
Oxygen(O2)Gas Sensor
-
Oxygen(O2)Gas Sensor
-
High Quality O2 Sensors