ORAKEL Chlorine Sensor
-
Price
Price on request ≥1piece
-
Model Number
MMS-D010 -
Data Sheet
-
Package
Original packaging -
Specs
MMS-D010 -
Quantity
- +Send Inquiry
Find Similar Products
-
Laboratory Dissolved Oxygen Sensor
Model Number:DO1200
MOQ:1
-
Submersible Galvanic Dissolved Oxygen Sensor
Model Number:DO6400/DO7400 Series
MOQ:1
-
UVT-LED Sensor UV Transmittance Sensor
Model Number:UVT-LED
MOQ:1
-
4-20mA Dissolved Oxygen Transmitters
Model Number:DOMA-10/DOMA-20
MOQ:1
-
Dissolved Oxygen Sensor for Oxygen Measurement
Model Number:SE 706
MOQ:1
-
Memosens Oxygen Sensor
Model Number:SE 715
MOQ:1
Product Description
ORAKEL CHLORINE SENSOR
The ORAKEL range of Residual Chlorine Analysers, Residual Chlorine Controllers and Residual Chlorine Monitors are membrane devices which are insensitive to changing pH, use no reagents, are extremely stable, and have reduced maintenance and reduced whole life costs.
The membraned amperometric chlorine sensors are enhanced with a third, reference electrode which eliminates zero drift. Its unique design means that pH correction is not usually required at all, completely eliminating reagents.
In addition to the state of the art potentiostatic chrono-amperometric free chlorine and total chlorine sensors, the ORAKEL range of residual chlorine analysers has all the functionality that you need, and more.
ORAKEL CHLORINE SENSOR APPLICATIONS
The ORAKEL Chlorine Sensor range is particularly suited to working in sites where reliability and ease of use are most important.
ORAKEL CHLORINE SENSOR TECHNICAL SPECIFICATION
The ORAKEL range of Residual Chlorine Analysers, Residual Chlorine Controllers and Residual Chlorine Monitors are membrane devices which are insensitive to changing pH, use no reagents, are extremely stable, and have reduced maintenance and reduced whole life costs.
The membraned amperometric chlorine sensors are enhanced with a third, reference electrode which eliminates zero drift. Its unique design means that pH correction is not usually required at all, completely eliminating reagents.
In addition to the state of the art potentiostatic chrono-amperometric free chlorine and total chlorine sensors, the ORAKEL range of residual chlorine analysers has all the functionality that you need, and more.
ORAKEL CHLORINE SENSOR APPLICATIONS
The ORAKEL Chlorine Sensor range is particularly suited to working in sites where reliability and ease of use are most important.
Chlorine Sensor Options
• Free Chlorine
• Total Chlorine
• Free Chlorine 'zero'
Water Treatment Applications
• Chlorine Dosing
• Paper Mills
• Cooling Towers
• Remote Sites
• Food Preparation
• Secondary Chlorination
ORAKEL CHLORINE SENSOR TECHNICAL SPECIFICATION
Type |
Membrane covered potentiostatic chronoamperometric three-electrode system |
Measurement |
Total chlorine or free chlorine |
Range |
0.01-2mg/l, 0.01-5mg/l, 0.01-10mg/l, 0.01-20mg/l and 0.5-200mg/l (free only). |
Resolution |
0.01mg/l (ppm) (0.1 on 0-200mg/l ranges). |
Reproducibility |
Better than ±0.05mg/l. |
Stability |
-1% per month (without calibration). Working electrode: gold. Counter electrode: stainless steel. Reference electrode: silver/silver halide |
Membrane Material |
Membrane Material. |
Flow Rate |
Approximately 0.5l/minute (minimum 0.25l/minute). |
Temperature Range |
0 to <45°C. |
Temperature Compensation |
Automatically by an integrated thermistor. |
pH Range |
pH 4 up to pH 9. |
First Polarisation Time |
2 hours. |
Re-Polarisation Time |
30 minutes. |
Response Time |
T90: approximately 120 seconds. |
Zero-Point Adjustment |
Not necessary. |
Calibration |
Manual using electrochemical test kit or DPD test kit. |
Housing Material |
PVC, silicone and polycarbonate stainless steel. |
Dimensions |
Diameter: approximately 25mm. Length: 175mm. |
Maintenance Intervals |
Membrane: once a year (depending on water quality). Electrolyte: once a year (depending on water quality). |
Interferences |
High levels of other oxidants such as Ozone and Chlorine Dioxide |