Electrochemical Hydrogen Cyanide Sensor (HCN Sensor) - HCN/C-100

  • Electrochemical Hydrogen Cyanide Sensor (HCN Sensor)
Mouse over image to zoom

• Nominal Range: 0 - 100 ppm
• Maximum Overload: n.d
• Expected Operation Life: 2 years in air
• Output Signal: 600 ± 150 nA/ppm
• Resolution: 0.2 ppm
• Repeatability: < 2 % of signal
• T90 Response Time: < 20 sec
• Temperature Range: -20°C to 50°C
• Relative Humidity Range: 15 % to 90 % R.H. non-condensing

  • Model Number: HCN/C-100
  • Data Sheet: Date Sheet File
Quantity:
- +

Product Specification

Compact Gas Sensors
Highest reliability and maximum performance in a compact housing. Suitable in most of all applications.

Features
Air vent on the back
O-ring on the front

Dimensions
Conforms to 7-Series
Diameter: 32 mm
Height: 17 mm
Pin diameter: 1.0 mm

HCN/C-100 Hydrogen Cyanide Sensor (HCN Sensor) Applications
• Safety and Environmental Control

HCN/C-100 Hydrogen Cyanide Sensor (HCN Sensor) Performance Characteristics
Performance data conditions: 20°C, 50% RH and 1013 mbar

Nominal Range
0 - 100 ppm
Maximum Overload
n.d
Expected Operation Life
2 years in air
Output Signal
600 ± 150 nA/ppm
Resolution
0.2 ppm
Temperature Range
-20°C to 50°C
Pressure Range
Atmospheric ± 10%
Pressure Coefficient
No data
T90 Response Time
< 20 sec
Relative Humidity Range
15 % to 90 % R.H. non-condensing
Typical Baseline Range (pure air, 20°C)
< 0.2 ppm
Maximum Zero Shift
n.d.
Long Term Output Drift
n.d.
Recommended Load Resistor
10 Ohm
Bias Voltage
Not required
Repeatability
< 2 % of signal
Output Linearity
Linear

HCN/C-100 Hydrogen Cyanide Sensor (HCN Sensor) Physical Characteristics
Weight
~ 13g
Position Sensitivity
None
Storage Life
Six months in container
Recommended Storage Temperature
5°C - 20°C
Warranty Period
12 months from date of dispatch

HCN/C-100 Hydrogen Cyanide Sensor (HCN Sensor) Cross Sensitivity Data
Interfering Gas
Concentration
Reading
CO
100 ppm
~ 2 ppm
H2S
15 ppm
~ 25 ppm
SO2 20 ppm ~ 38 ppm
NO
35 ppm
0 ppm
NO2
5 ppm
~ -12 ppm
H2
100 ppm
~ 2 ppm
Ethylene
100 ppm
0 ppm

If you cannot find what you want, you can entrust OFweek to source for you. Just click: