Dissolved Oxygen Sensor
-
Price
Price on request ≥1piece
-
Model Number
SE 707 -
Data Sheet
-
Package
Original packaging -
Specs
SE 707 -
Quantity
- +Send Inquiry
Find Similar Products
-
Laboratory Dissolved Oxygen Sensor
Model Number:DO1200
MOQ:1
-
Submersible Galvanic Dissolved Oxygen Sensor
Model Number:DO6400/DO7400 Series
MOQ:1
-
UVT-LED Sensor UV Transmittance Sensor
Model Number:UVT-LED
MOQ:1
-
4-20mA Dissolved Oxygen Transmitters
Model Number:DOMA-10/DOMA-20
MOQ:1
-
Dissolved Oxygen Sensor for Oxygen Measurement
Model Number:SE 706
MOQ:1
-
Memosens Oxygen Sensor
Model Number:SE 715
MOQ:1
Product Description
Reliable measurement of oxygen traces, also under sterile and hygienic conditions; signal-stable and robust
High resolution of low oxygen concentra-tions, simple and cost-efficient sensor cali- bration in ambient air. No calibration gases. High signal stability, also in the presence of high CO2 partial pressures.
FDA-conforming material with extremely smooth polished surface.
Simple main tenance due to modular design. Sterilizable, autoclavable, and CIP capable. Also suitable for hazardous locations.
SE 707 Dissolved Oxygen Sensor Applications
Beverage filling (e.g. milk, beer), measurement in boiler feed water (detection limit 1 ppb)
SE 707 Dissolved Oxygen Sensor Facts
- high level of process safety
- durable materials
- robust design
- easy to clean thanks to extremely smooth surface
- very simple calibration in air without calibration gases
- cost-efficient service due to easy-to-replace membrane module and interior body
- hygienic design
- very high resolution of 1 ppb in the trace range
- low cross-sensitivity to CO2
- operation also in hazardous locations
- VP screw cap
SE 707 Dissolved Oxygen Sensor Specifications
Range |
pO2 < 1200 mbar |
Detection limit |
1 ppb (3 ppb in liquids containing CO2 ) |
Accuracy |
1 % + 1 ppb (1 % + 3 ppb in liquids containing CO2 ) |
Response time at 25℃ (air -> N2 ) |
98 % full scale < 90 s |
Signal in ambient air |
290 ~ 500 nA |
Residual signal |
in O2 -free medium: ≤ 0.01 % of signal in ambient air in CO2: ≤ 0.03 % of signal in ambient air |
Flow dependence |
≤ 5 % |
Permissible pressure range (measurement) |
0.2 ~ 6 bar absolute |
Mechanical pressure resistance |
max. 12 bar absolute |
Permissible temperature range (measurement) |
0 ~ 80℃ |
Temperature range (stability) |
-5 ~ +121℃ (steam sterilizable/autoclavable) |
O-ring material |
silicone (FDA&USP Class VI tested) |
Membrane material |
PTFE/silicone/PTFE, FDA comp. (reinforced w. steel mesh) |
Process-wetted sensor parts |
1.4404 stainless steel (material certificate 3.1B) |
Surface roughness |
N 5 (RA < 0.4 μm) |
Temperature detector |
NTC 22 kohms |
Interior body incl. electrodes |
replaceable (ZU 0568) |
Sensor cap |
VP (VarioPin) |
ATEX Certificate |
II 1/2 GD Ex ia IIC T6/T5/T4/T3 |
Certificates |
3A, 3.1B (EN 10204/3.1.B) |