Further advantages of the HYT 939 digital humidity module include:
• Calibrated and temperature compensated
• Excellent humidity/temperature accuracy and stability
• High chemical resistance
• I2C protocol (address 0x28 or alternative address)
• Wide humidity and temperature range (0 % RH to 100 % RH and -40 °C to +125 °C)
• Very low drift
• Very stable at high humidity
• Interchangeable without adjustments
• Pressure-resistant version (up to 16 bar) upon request
The digital HYT 939 humidity module is suitable for various medical systems, autoclaves, pressure dew point measurement (pressure tight packaging required), drying systems and laboratory applications.
Technical Data:
Operating temperature range |
-40℃ to +125℃ |
|
Operating humidity range |
0 % RH to 100 % RH |
|
Hysteresis |
< ±1 % RH |
|
Linearity error |
< ±1 % RH |
|
Temperature error |
0.05 % RH/K (0℃ to +60℃) |
|
Operating voltage |
2.7 V to 5.5 V |
|
Current consumption (nominal) |
< 22 µA at 1 Hz measuring rate; 850 µA max. |
|
Current consumption (sleep) |
< 1 µA |
|
Digital interface |
I2C, address 0x28 or alternative address |
|
Operating voltage (limits) |
-0.3 V to 6 V |
|
Storage condition |
-20℃ to +50℃ |
|
|
Humidity |
Temperature |
Accuracy |
±1.8 % RH at +23℃ (0 % RH to 90 % RH) |
±0.2 K (0℃ to +60℃) |
Reproducibility |
±0.2 % RH |
±0.1 K |
Resolution |
0.02 % RH |
0.015℃ |
Response time t63 |
< 10 s with metal mesh filter |
< 10 s with metal mesh filter |
Long-term drift |
< 0.5 % RH/a (at +23℃ and 30 % RH to 70 % RH - laboratory conditions) |
< 0.05 K/a |
Measuring principle |
Capacitive polymer humidity sensor |
PTAT (integrated) |
Further advantages of the IST AG HYT LabKit include:
• Dew/frost point calculation and further humidity parameters
• Plug-in connection adapter included
• Evaluation software RECORDER and PCLOG for Windows included