4 Electrode Minature Dual Gas Sensor for CO and H2S - 4COSH CiTiceL
• Operating Principle: 3-electrode electrochemical
• Measurement Range:
CO: 0-500 ppm
H2S: 0-200 ppm
• Maximum Overload:
CO: 1500 ppm
H2S: 500 ppm
• Sensitivity:
CO: 80 ± 30 nA/ppm
H2S: 775 ± 275 nA/ppm
- Quantity:
- - +
Product Specification
4COSH CiTiceL
Carbon Monoxide / Hydrogen Sulfide (CO / H2S) Dual Gas Sensor
4COSH CiTiceL CO/H2S Dual Gas Sensor Key Features & Benefits:
• Robust, industry standard 4-Series packaging
• Compact Size
4COSH CiTiceL CO/H2S Dual Gas Sensor Technical Specifications
MEASUREMENT | |
Operating Principle | 3-electrode electrochemical |
Measurement Range |
CO: 0-500 ppm H2S: 0-200 ppm |
Maximum Overload |
CO: 1500 ppm H2S: 500 ppm |
Filter | To remove H2S from CO channel |
Sensitivity* |
CO: 80 ± 30 nA/ppm H2S: 775 ± 275 nA/ppm |
Response Time (T90)* |
CO: <35 Seconds at 20°C H2S: <35 Seconds at 20°C |
Typical Baseline Range |
CO: -2 to +3 ppm equiv. H2S: -0.4 to +0.4 ppm equiv. |
Resolution When using recommended electronics |
CO: 1.0 ppm H2S: 0.5 ppm |
Repeatability |
CO: < 3% of signal H2S: < 2% of signal |
Linearity | Linear |
ELECTRICAL | |
Recommended Load Resistor | 10 Ω |
Bias Voltage | Not required |
MECHANICAL | |
Weight | 5 g approx. |
Housing Material | ABS |
Orientation | Any |
ENVIRONMENTAL | |
Operating Temperature Range | -20°C to +50°C |
Recommended Storage Temp | 0°C to 20°C |
Operating Pressure Range | Atmospheric ± 10% |
Operating Humidity Range | 15 - 90% RH non-condensing |
LIFETIME | |
Long Term Sensitivity Drift | <30% signal loss/operating life |
Expected Operating Life | 3 years in air |
Storage Life | 6 months in CTL container |
* Specifications are valid at 20°C, 50% RH and 1013 mBar, using City Technology recommended circuitry. Performance characteristics outline the performance of sensors supplied within the first 3 months. Output signal can drift below the lower limit over time.
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Electrochemical carbon monoxide sensor CO
-
Carbon Monoxide Sensor (CO Sensor)
-
Hydrogen Cyanide Sensor (HCN Sensor)
-
Phosphine Sensor (PH3 Sensor)
-
Nitric Oxide Sensor (NO Sensor)
-
Nitrogen Dioxide Sensor (NO2 Sensor)
-
Infrared Carbon Dioxide (CO2) Sensor Pyroelectric Detector
-
Ethylene Oxide Sensor (ETO Sensor)
-
Individual Sensor Board (ISB) Alphasense B4 4-Electrode Gas Sensors
-
Hydrogen Chloride Sensor (HCL Sensor)