15x15 Filtered 0-5ppm NO2 Sensor C Package 110-508 - 3SP_NO2_5F C
• Measurement Range: 0 to 5 ppm
• Resolution: < 20 ppb
• Repeatability: < +/- 5 % of reading or 10ppb
• Response Time: < 15 seconds
• Sensitivity @ -25 mV bias: -30 +/- 15 nA/ppm
• Operating Temperature: -40 to 50°C
• Operating Humidity: 0 to >95% RH non-condensing
• Power Consumption: 10 to 50 uW
- Quantity:
- - +
Product Specification
3SP_NO2_5F C
15x15 Filtered 0-5 ppm NO2 Sensor in C Package
3SP_NO2_5F C Nitrogen Dioxide Sensor NO2 Sensor Description
SPEC’s printed sensors offer the performance of the best quality electrochemical sensors at a fraction of the price. SPEC’s printed sensors are also ultra-thin, offering easy integration into wireless, portable, and networked solutions. These sensors are ideal for health, environmental, industrial and residential monitoring, because of their high performance, low cost and small size.
3SP_NO2_5F C Nitrogen Dioxide Sensor NO2 Sensor Benefits
• NEW Onboard Filter Removes Ozone!!!
• Improved stability and low ppm sensitivity
• Small Size with Low Profile (20 x 20 x 3 mm)
• Long Life (10 years expected life)
• Fast Response (< 15 seconds)
• Low Power Consumption
• ROHS Compliant
3SP_NO2_5F C Nitrogen Dioxide Sensor NO2 Sensor Applications
• Air Quality Monitoring
• Industrial Safety
• Air Purification Control
3SP_NO2_5F C Nitrogen Dioxide Sensor NO2 Sensor Specifications
Measurement Range |
0 to 5 ppm |
Resolution |
< 20 ppb (instrumentation dependent) |
Resolution |
< 20 ppb (instrumentation dependent) |
Repeatability |
< +/- 5 % of reading or 10ppb |
Response Time |
< 15 seconds |
Sensitivity @ -25 mV bias |
-30 +/- 10 nA/ppm |
Expected Operating Life |
> 5 years (@ 23+/-3C; 40+/-10% RH) |
Operating Temperature Range |
-40 to 50°C (-20 to 40°C recommended) |
Operating Humidity Range - non-condensing |
0 to >95% RH (15 to 95% continuous) |
Power Consumption |
10 to 50 uW (circuit & ambient NO2 dependent) |
3SP_NO2_5F C Nitrogen Dioxide Sensor NO2 Sensor Cross Sensitivity
Most chemical sensors exhibit some cross-sensitivity to other gases. The following table lists the relative response of common potential interfering gases, and the concentration at which the data was gathered. Additional data will be made available as it is collected. Contact SPEC sensors for additional information.
Gas |
ppm |
Typical Response as ppm NO2 |
Carbon Monoxide |
100 |
<0.02 |
Chlorine |
10 |
<0.5 |
Ethanol |
100 |
<1 |
Hydrogen Sulfide |
5 |
<0.02 |
Nitric Oxide (NO) |
5 |
<0.1 |
Nitrogen Dioxide |
5 | 5 |
Ozone |
1 |
<0.1 |
Sulfur Dioxide |
5 |
<0.02 |
If you cannot find what you want, you can entrust OFweek to source for you. Just click:
Sourcing ServiceRecommended Products
-
Electrochemical carbon monoxide sensor CO
-
Carbon Monoxide Sensor (CO Sensor)
-
Hydrogen Cyanide Sensor (HCN Sensor)
-
Phosphine Sensor (PH3 Sensor)
-
Nitric Oxide Sensor (NO Sensor)
-
Nitrogen Dioxide Sensor (NO2 Sensor)
-
Infrared Carbon Dioxide (CO2) Sensor Pyroelectric Detector
-
Ethylene Oxide Sensor (ETO Sensor)
-
Individual Sensor Board (ISB) Alphasense B4 4-Electrode Gas Sensors
-
Hydrogen Chloride Sensor (HCL Sensor)